Printversion of this page
PDF-Version of this page

 
Printversion of this page
PDF-Version of this page
 
 

Optical Particle Counter OPC

Purpose of Participation

  • Airborne in-situ total aerosol number density of the accumulation mode aerosol > 0.35 µm in close to dry state
  • Ground based aerosol size distribution between 0.35 and 6.5 µm

Instrument description

The airborne OPC (GRIMM, model 7.309) delivers single particle counts and size classification based on aerosol scattering properties in a real time.
The ground-based OPC (GRIMM, model 1.106) delivers single particle counts and size classification based on aerosol scattering properties in a real time.


 
Species Aerosols
Method Direct concentration measurements
Quantity primary Aerosol number concentration/Aerosol size distribution
Quantity secondary  
Measurement range Particles between 0.35 and 6.5 µm up to 2000 cm -3
Altitude range NA
Vertical resolution NA
Temporal resolution 1 minute
Operation Cabin T > 0°C
Precision Depends on aerosol concentration, reproducibility ±3%
Preliminary data in principle next day (man power dependent)
Final data one year

 

Specific remarks

  • Ambient p and T from POLAR 4 basic instrumentation are required to derive ambient concentrations

References

GRIMM Aerosol Technik GmbH &Co.KG