OPC237H: Optical Particle Counter

Purpose of Participation


  • Air borne measurement of number concentration and size distribution of aerosol particles.


Instrument description

The instrument 237H counts and measures the scattered light by aerosol particles. The ranges of measurements are >0.1, 0.2, 0.3, 0.5, 0.7 and 1 m. Scattered light is detected by a 90° mini-HeNe sensor and converted to a pulse signal. An electric device measures the height of each pulse to obtain the information on the particle size. The raw data from the detector is also sent to Pulse Height Analyzer to obtain size distribution.


 
Species Aerosol (in situ)
Method Measurement of scattered light by aerosol particles
Quantity primary Number concentration of aerosol particle
Quantity secondary Size distribution of aerosol particles
Spectral range 0.1 - 1 m
Altitude range 0 - 8 km (up to aircraft limit)
Vertical resolution Several 10 m
Temporal resolution 2 seconds (highest) for OPC(6 size ranges) and 1 minute for PHA
Operation Continuously in any condition
Precision  
Data availability (during campaign) Next day
Final availability (after campaign) Three month

 
Printversion of this page
PDF-Version of this page
 

Contact Information

Head: Sadamu Yamagata (Hokkaido University)